19.6. Coincidence Timing
Table of Contents
Time difference
In PET, each coincidence event is built from two singles that are close in time. The time difference is simply the detection time of one single minus the detection time of the other. If the two detection times are $t_1$ and $t_2$, the time difference is usually defined as
$$
\Delta t = t_2 - t_1.
$$
GATE stores a time for every single event after digitization. When coincidences are sorted, the coincidence module or actor reads the times of the two singles and calculates this $\Delta t$. The sign of $\Delta t$ depends on which detector is considered first, so in many analyses you use $|\Delta t|$ rather than $\Delta t$ itself.
In non time of flight PET, the absolute value of $\Delta t$ is mainly used to decide if two singles belong to the same annihilation event. In time of flight PET, the value of $\Delta t$ also carries position information along the line of response, but that is addressed in the dedicated TOF chapter.
The time difference is affected by several factors: the true physical emission times of the photons, their different path lengths through the phantom and the scanner, and the detector timing response, including any timing blurring you have configured in the digitizer. In GATE, if you add a timing blurring component, each single time is shifted by a random delay that follows a chosen distribution, often Gaussian. This directly broadens the distribution of $\Delta t$.
In analysis, you often look at a histogram of $\Delta t$ values for all coincidences. For a stationary source and a realistic timing model this histogram has a narrow peak around zero. The width of this peak is related to the coincidence timing resolution that you have configured in your simulation.
The coincidence time difference is
$$\Delta t = t_2 - t_1.$$
A narrow distribution of $\Delta t$ around zero is essential for accurate coincidence selection and for realistic time of flight behavior.
Coincidence timing window
The coincidence timing window is the maximum allowed time difference between two singles for them to be accepted as a coincidence. In other words, two singles are considered coincident if
$$
|\Delta t| \le \Delta t_{\text{window}},
$$
where $\Delta t_{\text{window}}$ is the timing window half width that you choose.
In GATE, the coincidence sorter in the digitizer chain uses this rule. You configure the timing window as a parameter of the coincidence module, usually in units like ns. During the simulation, the sorter examines the ordered list of singles. When it finds two singles that come from different detectors and whose times satisfy this condition, it groups them into a coincidence event and writes them to the coincidence output.
The choice of $\Delta t_{\text{window}}$ has a strong impact on coincidence data quality. If the window is too narrow, many true coincidences are lost because detector timing blurring and different photon paths make their $|\Delta t|$ slightly larger than the ideal value. This reduces sensitivity and increases statistical noise. If the window is too wide, many unrelated singles that happen to occur close in time are grouped into coincidences, which increases random coincidences and degrades image quality.
For a given detector timing resolution, there is a practical range of useful timing windows. In a basic non time of flight PET simulation, the window is often chosen to be several times larger than the single detector timing resolution so that most true coincidences are accepted, while still rejecting a large fraction of randoms.
In time of flight PET, the same timing window also limits which events can be considered for TOF reconstruction. The detailed relation between $\Delta t_{\text{window}}$ and spatial localization along the line of response is described in the TOF PET chapter, but the basic coincidence selection rule is the same.
When you configure and analyze coincidences in GATE, it is useful to check the distribution of $\Delta t$ after the simulation. If the chosen timing window barely covers the central peak of that distribution, you are discarding many valid events. If the window is much wider than the central peak, you are allowing many unnecessary randoms. Adjusting $\Delta t_{\text{window}}$ is a simple but important part of realistic PET coincidence modeling.
Coincidence selection rule:
$$|\Delta t| \le \Delta t_{\text{window}}.$$
A smaller window reduces randoms but can lose true events. A larger window keeps more trues but increases random coincidences.
Views: 0
Where to Move
Move chapter:
- ☰ 1. Introduction to GATE
- ☰ 1.1. What Is GATE?
- ☰ 1.2. Applications of GATE
- ☰ 1.3. GATE vs Geant4
- ☰ 1.4. GATE Simulation Workflow
-
- ☰ 2. Installing GATE
- ☰ 2.1. GATE Requirements
- ☰ 2.2. Creating a Python Environment
- ☰ 2.3. Installing GATE
- ☰ 2.4. Verifying the Installation
- ☰ 2.5. GATE Example Simulations
-
- ☰ 3. Python Basics for GATE
- ☰ 3.1. Variables and Data Types
- ☰ 3.2. Conditional Statements
- ☰ 3.3. Loops
- ☰ 3.4. Functions
- ☰ 3.5. Python Modules
- ☰ 3.6. NumPy Basics
- ☰ 3.7. Working with Files
-
- ☰ 4. Creating Your First GATE Simulation
- ☰ 4.1. Importing GATE
- ☰ 4.2. Creating a Simulation
- ☰ 4.3. Adding the World
- ☰ 4.4. Adding a Particle Source
- ☰ 4.5. Adding an Actor
- ☰ 4.6. Running the Simulation
- ☰ 4.7. Inspecting the Results
-
- ☰ 5. Units in GATE
- ☰ 5.1. Physical Units
- ☰ 5.2. Length Units
- ☰ 5.3. Energy Units
- ☰ 5.4. Time Units
- ☰ 5.5. Radioactivity Units
- ☰ 5.6. Using Units Correctly
-
- ☰ 6. Geometry Fundamentals
- ☰ 6.1. The World Volume
- ☰ 6.2. Creating Volumes
- ☰ 6.3. Box Geometry
- ☰ 6.4. Cylindrical Geometry
- ☰ 6.5. Spherical Geometry
- ☰ 6.6. Position and Translation
- ☰ 6.7. Rotation
-
- ☰ 7. Building Complex Geometry
- ☰ 7.1. Nested Volumes
- ☰ 7.2. Repeated Volumes
- ☰ 7.3. Ring Geometry
- ☰ 7.4. Boolean Geometry
- ☰ 7.5. Geometry Overlap
- ☰ 7.6. Reusable Geometry Functions
-
- ☰ 8. Materials
- ☰ 8.1. Material Definitions
- ☰ 8.2. Geant4 Material Database
- ☰ 8.3. Common Medical Materials
- ☰ 8.4. Detector Materials
- ☰ 8.5. Shielding Materials
- ☰ 8.6. Creating Custom Materials
-
- ☰ 9. Geometry Visualization
- ☰ 9.1. Enabling Visualization
- ☰ 9.2. Displaying Volumes
- ☰ 9.3. Changing the View
- ☰ 9.4. Volume Appearance
- ☰ 9.5. Visualizing Particle Tracks
- ☰ 9.6. Debugging with Visualization
-
- ☰ 10. Particle Sources
- ☰ 10.1. Generic Sources
- ☰ 10.2. Particle Types
- ☰ 10.3. Source Position
- ☰ 10.4. Source Direction
- ☰ 10.5. Source Energy
- ☰ 10.6. Source Activity
-
- ☰ 11. Radioactive Sources
- ☰ 11.1. Radioactive Decay
- ☰ 11.2. Medical Radionuclides
- ☰ 11.3. Positron Emitters
- ☰ 11.4. Gamma Emitters
- ☰ 11.5. Activity Distribution
- ☰ 11.6. Time-Dependent Activity
-
- ☰ 12. Physics Lists
- ☰ 12.1. What Is a Physics List?
- ☰ 12.2. Electromagnetic Physics
- ☰ 12.3. Gamma Physics
- ☰ 12.4. Electron Physics
- ☰ 12.5. Positron Physics
- ☰ 12.6. Hadronic Physics
- ☰ 12.7. Choosing Physics Models
-
- ☰ 13. Production Cuts and Transport Parameters
- ☰ 13.1. What Are Production Cuts?
- ☰ 13.2. Range Cuts
- ☰ 13.3. Region-Based Cuts
- ☰ 13.4. Accuracy vs Performance
-
- ☰ 14. Actors
- ☰ 14.1. What Is an Actor?
- ☰ 14.2. Simulation Statistics Actor
- ☰ 14.3. Energy Deposition Actor
- ☰ 14.4. Dose Actor
- ☰ 14.5. Phase Space Actor
- ☰ 14.6. Particle Fluence
- ☰ 14.7. Actor Filters
-
- ☰ 15. Dose Calculation
- ☰ 15.1. Energy Deposition
- ☰ 15.2. Absorbed Dose
- ☰ 15.3. Creating a Dose Actor
- ☰ 15.4. Dose Images
- ☰ 15.5. Statistical Uncertainty
- ☰ 15.6. Dose Profiles
-
- ☰ 16. Hits and Detector Response
- ☰ 16.1. Particle Interactions in Detectors
- ☰ 16.2. Hits
- ☰ 16.3. Hit Collections
- ☰ 16.4. Hit Attributes
- ☰ 16.5. Hits vs Detector Signals
-
- ☰ 17. Digitizers
- ☰ 17.1. What Is Digitization?
- ☰ 17.2. Digitizer Chain
- ☰ 17.3. Energy Summation
- ☰ 17.4. Readout
- ☰ 17.5. Energy Blurring
- ☰ 17.6. Spatial Blurring
- ☰ 17.7. Energy Thresholds
- ☰ 17.8. Time Blurring
-
- ☰ 18. Singles
- ☰ 18.1. What Are Singles?
- ☰ 18.2. Creating Singles
- ☰ 18.3. Detector Identification
- ☰ 18.4. Energy Windows
- ☰ 18.5. Analyzing Singles
-
- ☰ 19. Coincidences
- ☰ 19.1. What Are Coincidences?
- ☰ 19.2. Coincidence Sorting
- ☰ 19.3. True Coincidences
- ☰ 19.4. Scattered Coincidences
- ☰ 19.5. Random Coincidences
- ☰ 19.6. Coincidence Timing
- ☰ 19.7. Coincidence Output
-
- ☰ 20. Data Output
- ☰ 20.1. Simulation Output
- ☰ 20.2. ROOT Output
- ☰ 20.3. Image Output
- ☰ 20.4. Text-Based Output
- ☰ 20.5. Organizing Simulation Results
-
- ☰ 21. GATE and ROOT
- ☰ 21.1. Reading GATE ROOT Files
- ☰ 21.2. Inspecting Output
- ☰ 21.3. Energy Histograms
- ☰ 21.4. Time Histograms
- ☰ 21.5. Detector Maps
- ☰ 21.6. Filtering Events
-
- ☰ 22. GATE and Python Data Analysis
- ☰ 22.1. Reading Simulation Output with Python
- ☰ 22.2. Energy Spectra
- ☰ 22.3. Position Distributions
- ☰ 22.4. Timing Analysis
- ☰ 22.5. Statistical Analysis
- ☰ 22.6. Visualization
-
- ☰ 23. Voxelized Geometry
- ☰ 23.1. Introduction to Voxelized Geometry
- ☰ 23.2. Image-Based Volumes
- ☰ 23.3. Material Mapping
- ☰ 23.4. CT-to-Material Conversion
- ☰ 23.5. Patient Geometry
- ☰ 23.6. Visualizing Voxelized Geometry
-
- ☰ 24. Image and DICOM Data
- ☰ 24.1. Medical Image Formats
- ☰ 24.2. Reading CT Images
- ☰ 24.3. DICOM CT Data
- ☰ 24.4. RT Structure Sets
- ☰ 24.5. RT Plans
- ☰ 24.6. Coordinate Systems
-
- ☰ 25. Motion and Time-Dependent Simulations
- ☰ 25.1. Time in GATE
- ☰ 25.2. Moving Sources
- ☰ 25.3. Moving Detectors
- ☰ 25.4. Respiratory Motion
- ☰ 25.5. Scanner Rotation
- ☰ 25.6. Dynamic Imaging
-
- ☰ 26. PET Simulation Fundamentals
- ☰ 26.1. PET Physics
- ☰ 26.2. PET Scanner Components
- ☰ 26.3. PET Detector Geometry
- ☰ 26.4. PET Source
- ☰ 26.5. PET Physics Configuration
- ☰ 26.6. PET Detector Response
-
- ☰ 27. Building a PET Scanner
- ☰ 27.1. Creating the Detector Ring
- ☰ 27.2. Creating Detector Blocks
- ☰ 27.3. Creating Detector Crystals
- ☰ 27.4. Assigning Detector IDs
- ☰ 27.5. Adding a PET Source
- ☰ 27.6. Configuring the Digitizer
- ☰ 27.7. Running the PET Simulation
-
- ☰ 28. PET Coincidence Analysis
- ☰ 28.1. Coincidence Events
- ☰ 28.2. Energy Selection
- ☰ 28.3. Coincidence Time Window
- ☰ 28.4. Lines of Response
- ☰ 28.5. True Events
- ☰ 28.6. Scatter Events
- ☰ 28.7. Random Events
- ☰ 28.8. PET Performance Metrics
-
- ☰ 29. Time-of-Flight PET
- ☰ 29.1. TOF PET Principles
- ☰ 29.2. Detector Timing Resolution
- ☰ 29.3. Time Difference
- ☰ 29.4. Timing Blurring
- ☰ 29.5. Analyzing TOF Events
-
- ☰ 30. SPECT Simulation Fundamentals
- ☰ 30.1. SPECT Physics
- ☰ 30.2. SPECT Scanner Components
- ☰ 30.3. SPECT Radionuclides
- ☰ 30.4. Gamma Camera Geometry
- ☰ 30.5. SPECT Source
-
- ☰ 31. Building a Gamma Camera
- ☰ 31.1. Detector Head
- ☰ 31.2. Collimator
- ☰ 31.3. Shielding
- ☰ 31.4. Energy Resolution
- ☰ 31.5. Energy Window
- ☰ 31.6. Projection Data
-
- ☰ 32. SPECT Acquisition
- ☰ 32.1. Rotating Gamma Camera
- ☰ 32.2. Projection Angles
- ☰ 32.3. Acquisition Time
- ☰ 32.4. Energy Windows
- ☰ 32.5. Projection Images
- ☰ 32.6. SPECT Data Analysis
-
- ☰ 33. CT and X-Ray Simulation
- ☰ 33.1. X-Ray Physics
- ☰ 33.2. X-Ray Source
- ☰ 33.3. CT Geometry
- ☰ 33.4. Detector Array
- ☰ 33.5. Rotating CT Geometry
- ☰ 33.6. CT Projection Data
-
- ☰ 34. Radiation Therapy Simulation
- ☰ 34.1. External Beam Radiotherapy
- ☰ 34.2. Photon Beams
- ☰ 34.3. Electron Beams
- ☰ 34.4. Dose Calculation
- ☰ 34.5. Beam Geometry
- ☰ 34.6. Dose Validation
-
- ☰ 35. Proton Therapy Simulation
- ☰ 35.1. Proton Physics
- ☰ 35.2. Proton Beam
- ☰ 35.3. Water Phantom
- ☰ 35.4. Bragg Peak
- ☰ 35.5. Spread-Out Bragg Peak
- ☰ 35.6. Proton Dose Analysis
-
- ☰ 36. Internal Dosimetry
- ☰ 36.1. Radioactive Sources in the Body
- ☰ 36.2. Voxelized Activity Maps
- ☰ 36.3. Energy Deposition
- ☰ 36.4. Dose Maps
- ☰ 36.5. Radionuclide Therapy
- ☰ 36.6. Dose Analysis
-
- ☰ 37. Optical Photon Simulation
- ☰ 37.1. Optical Physics
- ☰ 37.2. Scintillation
- ☰ 37.3. Optical Properties
- ☰ 37.4. Optical Surfaces
- ☰ 37.5. Photodetectors
- ☰ 37.6. Optical Photon Detection
-
- ☰ 38. Phase-Space Data
- ☰ 38.1. What Is a Phase Space?
- ☰ 38.2. Recording Phase-Space Data
- ☰ 38.3. Phase-Space Files
- ☰ 38.4. Phase-Space Sources
- ☰ 38.5. Applications
-
- ☰ 39. Random Numbers and Reproducibility
- ☰ 39.1. Monte Carlo Random Numbers
- ☰ 39.2. Random Seeds
- ☰ 39.3. Independent Simulations
- ☰ 39.4. Reproducible Research
-
- ☰ 40. Multithreading and Performance
- ☰ 40.1. Multithreaded Simulation
- ☰ 40.2. Selecting the Number of Threads
- ☰ 40.3. Simulation Speed
- ☰ 40.4. Optimizing Geometry
- ☰ 40.5. Optimizing Output
- ☰ 40.6. Large Simulations
-
- ☰ 41. Running GATE on HPC Systems
- ☰ 41.1. Why Use HPC?
- ☰ 41.2. Running GATE Without Visualization
- ☰ 41.3. Job Scripts
- ☰ 41.4. Running with Slurm
- ☰ 41.5. Running Multiple Simulations
- ☰ 41.6. Combining Results
-
- ☰ 42. Simulation Validation
- ☰ 42.1. Why Validate a GATE Simulation?
- ☰ 42.2. Geometry Validation
- ☰ 42.3. Physics Validation
- ☰ 42.4. Comparing with Analytical Models
- ☰ 42.5. Comparing with Experimental Data
- ☰ 42.6. Statistical Validation
-
- ☰ 43. Debugging GATE Simulations
- ☰ 43.1. Python Errors
- ☰ 43.2. Geometry Problems
- ☰ 43.3. Material Errors
- ☰ 43.4. Source Problems
- ☰ 43.5. Physics Problems
- ☰ 43.6. Actor Problems
- ☰ 43.7. Digitizer Problems
- ☰ 43.8. Performance Problems
-
- ☰ 44. Writing Better GATE Simulations
- ☰ 44.1. Organizing Simulation Scripts
- ☰ 44.2. Configuration Files
- ☰ 44.3. Avoiding Hard-Coded Values
- ☰ 44.4. Reusable Components
- ☰ 44.5. Output Organization
- ☰ 44.6. Documentation
- ☰ 44.7. Version Control
-
- ☰ 45. Practical Example: Gamma-Ray Detector
- ☰ 45.1. Project Overview
- ☰ 45.2. Creating the World
- ☰ 45.3. Creating the Detector
- ☰ 45.4. Creating a Gamma Source
- ☰ 45.5. Configuring Electromagnetic Physics
- ☰ 45.6. Recording Energy Deposition
- ☰ 45.7. Adding Energy Resolution
- ☰ 45.8. Creating an Energy Spectrum
- ☰ 45.9. Analyzing the Spectrum
-
- ☰ 46. Practical Example: Radiation Shielding
- ☰ 46.1. Project Overview
- ☰ 46.2. Creating the Source
- ☰ 46.3. Creating the Shield
- ☰ 46.4. Selecting Shielding Materials
- ☰ 46.5. Changing Shield Thickness
- ☰ 46.6. Recording Transmitted Photons
- ☰ 46.7. Calculating Transmission
- ☰ 46.8. Calculating Attenuation
- ☰ 46.9. Comparing with the Exponential Attenuation Law
-
- ☰ 47. Practical Example: PET Scanner
- ☰ 47.1. Project Overview
- ☰ 47.2. Creating the PET Ring
- ☰ 47.3. Creating Detector Modules
- ☰ 47.4. Creating Detector Crystals
- ☰ 47.5. Adding an F-18 Source
- ☰ 47.6. Configuring PET Physics
- ☰ 47.7. Recording Hits
- ☰ 47.8. Creating Singles
- ☰ 47.9. Applying Energy Blurring
- ☰ 47.10. Applying the PET Energy Window
- ☰ 47.11. Sorting Coincidences
- ☰ 47.12. Identifying True Coincidences
- ☰ 47.13. Identifying Scattered Coincidences
- ☰ 47.14. Creating Lines of Response
- ☰ 47.15. Saving PET Data
-
Comments
Please login to add a comment.
Don't have an account? Register now!