28.2. Energy Selection
Table of Contents
511 keV photopeak
In PET, almost all useful coincidence events come from the two 511 keV photons produced when a positron annihilates with an electron. When these photons are fully absorbed in a scintillation crystal, the measured energy is close to 511 keV and forms a characteristic peak in the detected energy spectrum, called the 511 keV photopeak.
In a GATE PET simulation, you usually observe this photopeak in either singles or coincidence energy spectra after digitization. The raw energy deposition in a crystal is first summed over hits, then blurred by the detector energy resolution. Because of this finite resolution, the photopeak is not a single line at 511 keV but a broadened peak with a width that is often characterized by the full width at half maximum, FWHM. For a detector with energy resolution $R$ at 511 keV, you often express it as a percentage:
$$
R = \frac{\text{FWHM}}{E_0} \times 100\%
$$
with $E_0 = 511 \,\text{keV}$.
For PET energy selection, the 511 keV photopeak is the primary feature of interest. You must always take detector energy resolution into account, because the measured peak is broadened and not exactly at 511 keV for every event.
In simulated spectra, you may also see a continuum below 511 keV from Compton scattered photons, both those that scatter in the patient or phantom and those that scatter inside the detector before being absorbed. These events deposit less than 511 keV and populate lower energies. Identifying the photopeak means locating the local maximum around 511 keV and understanding how much of the spectrum below it is scatter.
In practice, you often begin by plotting a histogram of detected energies from singles or coincidence data and visually identify the photopeak region. Because you control the detector model and blurring in GATE, you can predict approximately where the peak should appear and what width it should have given your specified resolution.
PET energy windows
Energy selection in PET is implemented by defining an energy window around the 511 keV photopeak. The goal is to keep as many true, unscattered annihilation photons as possible and reject as many scattered photons as possible. This selection is applied to the energy of singles before forming coincidences, or to the energies in coincidence events, depending on your digitizer chain.
An energy window is usually defined by a lower and an upper energy threshold. For a nominal 511 keV photopeak, common clinical-like settings are for example 350 to 650 keV, or 400 to 600 keV. The lower threshold removes most scattered photons that lost a significant amount of energy, and the upper threshold removes events with abnormally high reconstructed energy, for example from pile up or multiple interactions summed together.
In GATE you typically configure the PET energy window as part of the digitizer. Singles that do not pass the energy window are discarded and do not contribute to coincidence formation. This has a direct impact on the number of random, scatter, and even true coincidences. A narrower window reduces scatter more effectively but also reduces sensitivity because some true events fall into the tails of the photopeak and are rejected.
You can think of the trade off in terms of accepted fraction of true events versus rejected scatter. If $N_\text{true}$ is the number of true events before energy selection and $N_\text{true,win}$ is the number that fall inside the window, then the true acceptance is
$$
f_\text{true} = \frac{N_\text{true,win}}{N_\text{true}}
$$
Similarly, for scatter events $N_\text{scat}$ and $N_\text{scat,win}$ the scatter acceptance is
$$
f_\text{scat} = \frac{N_\text{scat,win}}{N_\text{scat}}
$$
Ideal energy selection would give a high $f_\text{true}$ and a low $f_\text{scat}$, but in practice both depend on detector resolution and on how much scattering occurs in the object.
When choosing PET energy windows, always:
- Place the window around the 511 keV photopeak.
- Set a lower threshold high enough to reject most scattered photons.
- Avoid an excessively narrow window that removes too many true events, especially with poor energy resolution.
In a PET coincidence analysis workflow, energy selection is typically applied early. After running a simulation and producing singles or coincidence ROOT files, you can reproduce the same energy window in analysis by selecting events with energies between $E_\text{low}$ and $E_\text{high}$. This allows you to study how different energy windows affect sensitivity, scatter fraction, and count rates without rerunning the full simulation, as long as you initially saved events over a broad energy range.
As you explore PET performance metrics, it is useful to systematically vary the PET energy window and measure how the number of true, scattered, and random coincidences changes. In a GATE based study, you can quantify how tighter energy selection improves image quality indicators, at the cost of reduced statistics and longer acquisition or simulation times needed to achieve a given level of noise.
Views: 0
Where to Move
Move chapter:
- ☰ 1. Introduction to GATE
- ☰ 1.1. What Is GATE?
- ☰ 1.2. Applications of GATE
- ☰ 1.3. GATE vs Geant4
- ☰ 1.4. GATE Simulation Workflow
-
- ☰ 2. Installing GATE
- ☰ 2.1. GATE Requirements
- ☰ 2.2. Creating a Python Environment
- ☰ 2.3. Installing GATE
- ☰ 2.4. Verifying the Installation
- ☰ 2.5. GATE Example Simulations
-
- ☰ 3. Python Basics for GATE
- ☰ 3.1. Variables and Data Types
- ☰ 3.2. Conditional Statements
- ☰ 3.3. Loops
- ☰ 3.4. Functions
- ☰ 3.5. Python Modules
- ☰ 3.6. NumPy Basics
- ☰ 3.7. Working with Files
-
- ☰ 4. Creating Your First GATE Simulation
- ☰ 4.1. Importing GATE
- ☰ 4.2. Creating a Simulation
- ☰ 4.3. Adding the World
- ☰ 4.4. Adding a Particle Source
- ☰ 4.5. Adding an Actor
- ☰ 4.6. Running the Simulation
- ☰ 4.7. Inspecting the Results
-
- ☰ 5. Units in GATE
- ☰ 5.1. Physical Units
- ☰ 5.2. Length Units
- ☰ 5.3. Energy Units
- ☰ 5.4. Time Units
- ☰ 5.5. Radioactivity Units
- ☰ 5.6. Using Units Correctly
-
- ☰ 6. Geometry Fundamentals
- ☰ 6.1. The World Volume
- ☰ 6.2. Creating Volumes
- ☰ 6.3. Box Geometry
- ☰ 6.4. Cylindrical Geometry
- ☰ 6.5. Spherical Geometry
- ☰ 6.6. Position and Translation
- ☰ 6.7. Rotation
-
- ☰ 7. Building Complex Geometry
- ☰ 7.1. Nested Volumes
- ☰ 7.2. Repeated Volumes
- ☰ 7.3. Ring Geometry
- ☰ 7.4. Boolean Geometry
- ☰ 7.5. Geometry Overlap
- ☰ 7.6. Reusable Geometry Functions
-
- ☰ 8. Materials
- ☰ 8.1. Material Definitions
- ☰ 8.2. Geant4 Material Database
- ☰ 8.3. Common Medical Materials
- ☰ 8.4. Detector Materials
- ☰ 8.5. Shielding Materials
- ☰ 8.6. Creating Custom Materials
-
- ☰ 9. Geometry Visualization
- ☰ 9.1. Enabling Visualization
- ☰ 9.2. Displaying Volumes
- ☰ 9.3. Changing the View
- ☰ 9.4. Volume Appearance
- ☰ 9.5. Visualizing Particle Tracks
- ☰ 9.6. Debugging with Visualization
-
- ☰ 10. Particle Sources
- ☰ 10.1. Generic Sources
- ☰ 10.2. Particle Types
- ☰ 10.3. Source Position
- ☰ 10.4. Source Direction
- ☰ 10.5. Source Energy
- ☰ 10.6. Source Activity
-
- ☰ 11. Radioactive Sources
- ☰ 11.1. Radioactive Decay
- ☰ 11.2. Medical Radionuclides
- ☰ 11.3. Positron Emitters
- ☰ 11.4. Gamma Emitters
- ☰ 11.5. Activity Distribution
- ☰ 11.6. Time-Dependent Activity
-
- ☰ 12. Physics Lists
- ☰ 12.1. What Is a Physics List?
- ☰ 12.2. Electromagnetic Physics
- ☰ 12.3. Gamma Physics
- ☰ 12.4. Electron Physics
- ☰ 12.5. Positron Physics
- ☰ 12.6. Hadronic Physics
- ☰ 12.7. Choosing Physics Models
-
- ☰ 13. Production Cuts and Transport Parameters
- ☰ 13.1. What Are Production Cuts?
- ☰ 13.2. Range Cuts
- ☰ 13.3. Region-Based Cuts
- ☰ 13.4. Accuracy vs Performance
-
- ☰ 14. Actors
- ☰ 14.1. What Is an Actor?
- ☰ 14.2. Simulation Statistics Actor
- ☰ 14.3. Energy Deposition Actor
- ☰ 14.4. Dose Actor
- ☰ 14.5. Phase Space Actor
- ☰ 14.6. Particle Fluence
- ☰ 14.7. Actor Filters
-
- ☰ 15. Dose Calculation
- ☰ 15.1. Energy Deposition
- ☰ 15.2. Absorbed Dose
- ☰ 15.3. Creating a Dose Actor
- ☰ 15.4. Dose Images
- ☰ 15.5. Statistical Uncertainty
- ☰ 15.6. Dose Profiles
-
- ☰ 16. Hits and Detector Response
- ☰ 16.1. Particle Interactions in Detectors
- ☰ 16.2. Hits
- ☰ 16.3. Hit Collections
- ☰ 16.4. Hit Attributes
- ☰ 16.5. Hits vs Detector Signals
-
- ☰ 17. Digitizers
- ☰ 17.1. What Is Digitization?
- ☰ 17.2. Digitizer Chain
- ☰ 17.3. Energy Summation
- ☰ 17.4. Readout
- ☰ 17.5. Energy Blurring
- ☰ 17.6. Spatial Blurring
- ☰ 17.7. Energy Thresholds
- ☰ 17.8. Time Blurring
-
- ☰ 18. Singles
- ☰ 18.1. What Are Singles?
- ☰ 18.2. Creating Singles
- ☰ 18.3. Detector Identification
- ☰ 18.4. Energy Windows
- ☰ 18.5. Analyzing Singles
-
- ☰ 19. Coincidences
- ☰ 19.1. What Are Coincidences?
- ☰ 19.2. Coincidence Sorting
- ☰ 19.3. True Coincidences
- ☰ 19.4. Scattered Coincidences
- ☰ 19.5. Random Coincidences
- ☰ 19.6. Coincidence Timing
- ☰ 19.7. Coincidence Output
-
- ☰ 20. Data Output
- ☰ 20.1. Simulation Output
- ☰ 20.2. ROOT Output
- ☰ 20.3. Image Output
- ☰ 20.4. Text-Based Output
- ☰ 20.5. Organizing Simulation Results
-
- ☰ 21. GATE and ROOT
- ☰ 21.1. Reading GATE ROOT Files
- ☰ 21.2. Inspecting Output
- ☰ 21.3. Energy Histograms
- ☰ 21.4. Time Histograms
- ☰ 21.5. Detector Maps
- ☰ 21.6. Filtering Events
-
- ☰ 22. GATE and Python Data Analysis
- ☰ 22.1. Reading Simulation Output with Python
- ☰ 22.2. Energy Spectra
- ☰ 22.3. Position Distributions
- ☰ 22.4. Timing Analysis
- ☰ 22.5. Statistical Analysis
- ☰ 22.6. Visualization
-
- ☰ 23. Voxelized Geometry
- ☰ 23.1. Introduction to Voxelized Geometry
- ☰ 23.2. Image-Based Volumes
- ☰ 23.3. Material Mapping
- ☰ 23.4. CT-to-Material Conversion
- ☰ 23.5. Patient Geometry
- ☰ 23.6. Visualizing Voxelized Geometry
-
- ☰ 24. Image and DICOM Data
- ☰ 24.1. Medical Image Formats
- ☰ 24.2. Reading CT Images
- ☰ 24.3. DICOM CT Data
- ☰ 24.4. RT Structure Sets
- ☰ 24.5. RT Plans
- ☰ 24.6. Coordinate Systems
-
- ☰ 25. Motion and Time-Dependent Simulations
- ☰ 25.1. Time in GATE
- ☰ 25.2. Moving Sources
- ☰ 25.3. Moving Detectors
- ☰ 25.4. Respiratory Motion
- ☰ 25.5. Scanner Rotation
- ☰ 25.6. Dynamic Imaging
-
- ☰ 26. PET Simulation Fundamentals
- ☰ 26.1. PET Physics
- ☰ 26.2. PET Scanner Components
- ☰ 26.3. PET Detector Geometry
- ☰ 26.4. PET Source
- ☰ 26.5. PET Physics Configuration
- ☰ 26.6. PET Detector Response
-
- ☰ 27. Building a PET Scanner
- ☰ 27.1. Creating the Detector Ring
- ☰ 27.2. Creating Detector Blocks
- ☰ 27.3. Creating Detector Crystals
- ☰ 27.4. Assigning Detector IDs
- ☰ 27.5. Adding a PET Source
- ☰ 27.6. Configuring the Digitizer
- ☰ 27.7. Running the PET Simulation
-
- ☰ 28. PET Coincidence Analysis
- ☰ 28.1. Coincidence Events
- ☰ 28.2. Energy Selection
- ☰ 28.3. Coincidence Time Window
- ☰ 28.4. Lines of Response
- ☰ 28.5. True Events
- ☰ 28.6. Scatter Events
- ☰ 28.7. Random Events
- ☰ 28.8. PET Performance Metrics
-
- ☰ 29. Time-of-Flight PET
- ☰ 29.1. TOF PET Principles
- ☰ 29.2. Detector Timing Resolution
- ☰ 29.3. Time Difference
- ☰ 29.4. Timing Blurring
- ☰ 29.5. Analyzing TOF Events
-
- ☰ 30. SPECT Simulation Fundamentals
- ☰ 30.1. SPECT Physics
- ☰ 30.2. SPECT Scanner Components
- ☰ 30.3. SPECT Radionuclides
- ☰ 30.4. Gamma Camera Geometry
- ☰ 30.5. SPECT Source
-
- ☰ 31. Building a Gamma Camera
- ☰ 31.1. Detector Head
- ☰ 31.2. Collimator
- ☰ 31.3. Shielding
- ☰ 31.4. Energy Resolution
- ☰ 31.5. Energy Window
- ☰ 31.6. Projection Data
-
- ☰ 32. SPECT Acquisition
- ☰ 32.1. Rotating Gamma Camera
- ☰ 32.2. Projection Angles
- ☰ 32.3. Acquisition Time
- ☰ 32.4. Energy Windows
- ☰ 32.5. Projection Images
- ☰ 32.6. SPECT Data Analysis
-
- ☰ 33. CT and X-Ray Simulation
- ☰ 33.1. X-Ray Physics
- ☰ 33.2. X-Ray Source
- ☰ 33.3. CT Geometry
- ☰ 33.4. Detector Array
- ☰ 33.5. Rotating CT Geometry
- ☰ 33.6. CT Projection Data
-
- ☰ 34. Radiation Therapy Simulation
- ☰ 34.1. External Beam Radiotherapy
- ☰ 34.2. Photon Beams
- ☰ 34.3. Electron Beams
- ☰ 34.4. Dose Calculation
- ☰ 34.5. Beam Geometry
- ☰ 34.6. Dose Validation
-
- ☰ 35. Proton Therapy Simulation
- ☰ 35.1. Proton Physics
- ☰ 35.2. Proton Beam
- ☰ 35.3. Water Phantom
- ☰ 35.4. Bragg Peak
- ☰ 35.5. Spread-Out Bragg Peak
- ☰ 35.6. Proton Dose Analysis
-
- ☰ 36. Internal Dosimetry
- ☰ 36.1. Radioactive Sources in the Body
- ☰ 36.2. Voxelized Activity Maps
- ☰ 36.3. Energy Deposition
- ☰ 36.4. Dose Maps
- ☰ 36.5. Radionuclide Therapy
- ☰ 36.6. Dose Analysis
-
- ☰ 37. Optical Photon Simulation
- ☰ 37.1. Optical Physics
- ☰ 37.2. Scintillation
- ☰ 37.3. Optical Properties
- ☰ 37.4. Optical Surfaces
- ☰ 37.5. Photodetectors
- ☰ 37.6. Optical Photon Detection
-
- ☰ 38. Phase-Space Data
- ☰ 38.1. What Is a Phase Space?
- ☰ 38.2. Recording Phase-Space Data
- ☰ 38.3. Phase-Space Files
- ☰ 38.4. Phase-Space Sources
- ☰ 38.5. Applications
-
- ☰ 39. Random Numbers and Reproducibility
- ☰ 39.1. Monte Carlo Random Numbers
- ☰ 39.2. Random Seeds
- ☰ 39.3. Independent Simulations
- ☰ 39.4. Reproducible Research
-
- ☰ 40. Multithreading and Performance
- ☰ 40.1. Multithreaded Simulation
- ☰ 40.2. Selecting the Number of Threads
- ☰ 40.3. Simulation Speed
- ☰ 40.4. Optimizing Geometry
- ☰ 40.5. Optimizing Output
- ☰ 40.6. Large Simulations
-
- ☰ 41. Running GATE on HPC Systems
- ☰ 41.1. Why Use HPC?
- ☰ 41.2. Running GATE Without Visualization
- ☰ 41.3. Job Scripts
- ☰ 41.4. Running with Slurm
- ☰ 41.5. Running Multiple Simulations
- ☰ 41.6. Combining Results
-
- ☰ 42. Simulation Validation
- ☰ 42.1. Why Validate a GATE Simulation?
- ☰ 42.2. Geometry Validation
- ☰ 42.3. Physics Validation
- ☰ 42.4. Comparing with Analytical Models
- ☰ 42.5. Comparing with Experimental Data
- ☰ 42.6. Statistical Validation
-
- ☰ 43. Debugging GATE Simulations
- ☰ 43.1. Python Errors
- ☰ 43.2. Geometry Problems
- ☰ 43.3. Material Errors
- ☰ 43.4. Source Problems
- ☰ 43.5. Physics Problems
- ☰ 43.6. Actor Problems
- ☰ 43.7. Digitizer Problems
- ☰ 43.8. Performance Problems
-
- ☰ 44. Writing Better GATE Simulations
- ☰ 44.1. Organizing Simulation Scripts
- ☰ 44.2. Configuration Files
- ☰ 44.3. Avoiding Hard-Coded Values
- ☰ 44.4. Reusable Components
- ☰ 44.5. Output Organization
- ☰ 44.6. Documentation
- ☰ 44.7. Version Control
-
- ☰ 45. Practical Example: Gamma-Ray Detector
- ☰ 45.1. Project Overview
- ☰ 45.2. Creating the World
- ☰ 45.3. Creating the Detector
- ☰ 45.4. Creating a Gamma Source
- ☰ 45.5. Configuring Electromagnetic Physics
- ☰ 45.6. Recording Energy Deposition
- ☰ 45.7. Adding Energy Resolution
- ☰ 45.8. Creating an Energy Spectrum
- ☰ 45.9. Analyzing the Spectrum
-
- ☰ 46. Practical Example: Radiation Shielding
- ☰ 46.1. Project Overview
- ☰ 46.2. Creating the Source
- ☰ 46.3. Creating the Shield
- ☰ 46.4. Selecting Shielding Materials
- ☰ 46.5. Changing Shield Thickness
- ☰ 46.6. Recording Transmitted Photons
- ☰ 46.7. Calculating Transmission
- ☰ 46.8. Calculating Attenuation
- ☰ 46.9. Comparing with the Exponential Attenuation Law
-
- ☰ 47. Practical Example: PET Scanner
- ☰ 47.1. Project Overview
- ☰ 47.2. Creating the PET Ring
- ☰ 47.3. Creating Detector Modules
- ☰ 47.4. Creating Detector Crystals
- ☰ 47.5. Adding an F-18 Source
- ☰ 47.6. Configuring PET Physics
- ☰ 47.7. Recording Hits
- ☰ 47.8. Creating Singles
- ☰ 47.9. Applying Energy Blurring
- ☰ 47.10. Applying the PET Energy Window
- ☰ 47.11. Sorting Coincidences
- ☰ 47.12. Identifying True Coincidences
- ☰ 47.13. Identifying Scattered Coincidences
- ☰ 47.14. Creating Lines of Response
- ☰ 47.15. Saving PET Data
-
Comments
Please login to add a comment.
Don't have an account? Register now!